AC Power Supply PCR-LE series

AC Power Supply PCR-LE2 series

AC Power Supply PCR-M series

Commnunication Interface Mannual


 

Default Settings

Conditions after sending a *RST, fctory shipment, and at power-on

The following table shows how the PCR-LE2 is set when the *RST command is executed, at the time of factory shipment, and when the power is turned on.

 

Setup Item Setting Unit
*RST/ Factory default At poer-on
PCR
6000LE2
PCR
9000LE2
PCR
12000LE2
PCR
18000LE2
PCR
27000LE2
OUTP:COUP AC Setting immediately before turning the POWER switch off --
VOLT:RANG 150 --
VOLT 0 V
VOLT:OFFS 0 V
VOLT:LINE 0 V
VOLT:OFFS:LINE 0 V
FREQ 50 HZ
OUTP:PHAS:ON 0/OFF --
OUTP:PHAS:ON:LEV 0 DEG
OUTP:PHAS:OFF 0/OFF --
OUTP:PHAS:OFF:LEV 0 DEG
OUTP 0/OFF Depends on the OUTP:PON settins --
OUTP:PON SAFE --
FREQ:LIM 999.9 Setting immediately before turning the POWER switch off HZ
FREQ:LIM:LOW 1 HZ
CURR Single-phase 66 99 132 198 297 A
CURR
Single-phase, three-wire or Three-wire (each phase)
2233446699 A
CURR:OFFS Single-phase 46.269.392.4138.6207.9 A
CURR:OFFS
Single-phase, three-wire or Three-wire (each phase)
15.423.130.846.269.3 A
CURR:PROT:STAT 1/ON --
CURR:PROT:TRIP:TIME 10 S
CURR:PROT:OCP2:TIME 3 S
CURR:PEAK Single-phase 2643965287921188 A
CURR:PEAK
Single-phase, three-wire or Three-wire (each phase)
88132176264396 A
CURR:PEAK:LOWSingle-phase -264-396-528-792-1188 A
CURR:PEAK:LOW
Single-phase, three-wire or Three-wire (each phase)
-88-132-176-264 -396 A
VOLT:LIM 305.0 V
VOLT:LIM:LOW 0 V
VOLT:OFFS:LIM 431.0 V
VOLT:OFFS:LIM:LOW -431.0 V
VOLT:OFFS:PROT:LEV 474.1 V
VOLT:OFFS:PROT:LEV:LOW -474.1 V
VOLT:PROT:LEV 335.5 V
VOLT:PROT:LEV:LOW 0 V
SYST:CONF:RESP MED --
FREQ:SYNC 0/OFF --
FREQ:SYNC:PHASE:DEL 0 DEG
SYST:CONF:SENS:MODE 0/OFF --
WAVE:BANK 0 --
OUTP:IMP 0/OFF --
OUTP:IMP:LEV
Single-phase, L range
0.003330.00222 0.001670.00111 0.000741 OHM
OUTP:IMP:LEV
Single-phase, H range
0.01333 0.00889 0.00667 0.004440.00296 OHM
OUTP:IMP:LEV
Single-phase, three-wire or Three-wire (each phase), L range
0.010.006670.005 0.003330.00222 OHM
OUTP:IMP:LEV
Single-phase, three-wire or Three-wire (each phase), H range
0.04 0.02667 0.02 0.01333 0.00889 OHM
OUTP:IMP:RAT 1 PCT
OUTP:SST 0/OFF --
OUTP:SST:TIME 0.1 S
VOLT:VCC:MODE AUTO --
VOLT:VCC 431.0 V
DISP:MMOD:CURR RMS --
DISP:MMOD:VOLT RMS --
DISP:LVOL 0 --
DISP:PHAS 0 --
VOLT:TRIG 0 V
VOLT:OFFS:TRIG 0 V
VOLT:OFFS:MODE FIX --
VOLT:MODE FIX --
FREQ:TRIG 50.00 Setting immediately before turning the POWER switch off HZ
FREQ:MODE FIX --
WAVE:BANK:TRIG 0 Setting immediately before turning the POWER switch off --
WACE:BANK:MODE FIX --
TRIG:SOUR BUS --
TRIG:SEQ3:SOUR
TRIG:ACQ:SOUR
IMM --
SENS:APER 0.4 Setting immediately before turning the POWER switch off S
SENS:CURR:HOLD 0 S
SIM:T1:PHAS:STAT 0/OFF --
SIM:T1:TIME 0.1 S
SIM:T1:PHAS 0 DEG
SIM:T2:TIME 0 S
SIM:T3:TIME 0.1 S
SIM:T3:VOLT 0 V
SIM:T4:TIME 0 S
SIM:T5:CYCL:STAT 0/OFF --
SIM:T5:TIME 0.1 S
SIM:T5:CYCL 0 --
SIM:POL NORM --
SIM:REP:COUN 9999 --
TRIG:SEQ4:SOUR
TRIG:SIM:SOUR
IMM --
TRIG:SEQ5:SOUR
TRIG:PROG:SOUR
IMM --
SYST:SLE 0/OFF Setting immediately before turning the POWER switch off --
SYST:SLE:TIME 3600 S
SYST:KLOC 0/OFF --
FUNC:SOUR INT --
SYST:CONF:EXT:SINP:POL NORM --
VOLT:EXT:SOUR NONE --

>top

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

AC power supply  PCR-M seriesAC Power Supply PCR-LE series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-LE2 series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-M series
Communication Interface Manual