VOLT:LINE
Sets the AC line voltage.
This command is valid in AC mode (when the parameter to OUTP:COUP is AC).
This command is valid during single-phase, three-wire output (PCR-LE optional) when the U-V phase difference is 180 (SYST:CONF:PHAS:UV 180).
This command is valid during three-phase output (PCR-LE optional) when the U-V phase difference is 120 (SYST:CONF:PHAS:UV 120) and the U-W phase difference is 240 (SYST:CONF:PHAS:UW 240).
If the EX05-PCR-LE is in use (optional) and the signal source is set to external signal source (FUNC:SOUR EXT), this command is invalid.
If the EX06-PCR-LE is in use (optional) and the AC voltage is being controlled with an external signal (VOLT:EXT:SOUR VOLT), this command is invalid.
Command
[SOURce:]VOLTage:LINE {<numeric>|MINimum|MAXimum}
[SOURce:]VOLTage:LINE? [{MINimum|MAXimum}]
Parameter
Value | Single-phase, three-wire output L range: 0 to 305.0 H range: 0 to 610.0 Three-phase output L range: 0 to 264.1 H range: 0 to 528.2 A SCPI error (-222, "Data out of range") occurs if outside the range. The default value is 0. |
Unit | V |
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
Response
Response
Returns the AC line voltage setting in NR3 format in response to VOLT:LINE?.