VOLT:LINE

Sets the AC line voltage.

This command is valid in AC mode (when the parameter to OUTP:COUP is AC).

This command is valid during single-phase, three-wire output (PCR-LE optional) when the U-V phase difference is 180 (SYST:CONF:PHAS:UV 180).

This command is valid during three-phase output (PCR-LE optional) when the U-V phase difference is 120 (SYST:CONF:PHAS:UV 120) and the U-W phase difference is 240 (SYST:CONF:PHAS:UW 240).

If the EX05-PCR-LE is in use (optional) and the signal source is set to external signal source (FUNC:SOUR EXT), this command is invalid.

If the EX06-PCR-LE is in use (optional) and the AC voltage is being controlled with an external signal (VOLT:EXT:SOUR VOLT), this command is invalid.

Command

[SOURce:]VOLTage:LINE {<numeric>|MINimum|MAXimum}

[SOURce:]VOLTage:LINE? [{MINimum|MAXimum}]

Parameter

Value Single-phase, three-wire output
L range: 0 to 305.0
H range: 0 to 610.0
Three-phase output
L range: 0 to 264.1
H range: 0 to 528.2
 
A SCPI error (-222, "Data out of range") occurs if outside the range.
The default value is 0.
Unit V

For the setting that is applied when *RST is sent, see this table

For the setting that is applied when *RST is sent, see this table

For the setting that is applied when *RST is sent, see this table

Response

Response

Returns the AC line voltage setting in NR3 format in response to VOLT:LINE?.

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

AC power supply  PCR-M seriesAC Power Supply PCR-LE series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-LE2 series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-M series
Communication Interface Manual