Processing Time of Commands
The command processing time is the time until the next command is accepted.
The processing times indicated here are typical values.
They are not warranted.
The processing times vary depending on the settings and the measurement conditions.
It does not include the response time of the hardware.
PCR-LE/ PCR-LE2 series
Command | GPIB processing time*1 (ms) |
USB processing time (ms) |
RS232C processing time*2 (ms) |
LAN processing tim*3 (ms) |
Description |
---|---|---|---|---|---|
VOLT | 6 | 5 | 6 | 6 | Sets the voltage |
MEAS:VOLT? | 111 | 110 | 120 | 120 | Queries the measured voltage output |
FREQ | 6 | 6 | 6 | 6 | Sets the frequency |
MEAS:CURR? | 111 | 110 | 120 | 120 | Queries the measured voltage output |
*RST | 50 | 50 | 50 | 52 | Performs a device reset |
*1:Using GPIB-USB-B by National Instruments.
*2:Data rate setting: 19200 bps, Flow control: RTS・CTS
*3:100BASE-TX Ethernet