Command (Function)
Measured Value Queries
The second level nodes are SEQuence3 and ACQuire. Regardless of which node you use, the PCR operates in the same way when it receives the commands.
If you use the root node MEASure, the measurement is performed, and then the measured value is queried. If you use FETC, the measured value is queried without a measurement being performed.
TRIG:SEQ3:SOUR TRIG:ACQ:SOUR |
Sets the trigger source of the sequence 3 group |
---|---|
INIT:SEQ3 | Starts the trigger function of the sequence 3 group |
INIT:NAME | Starts the trigger function of the specified sequence group |
TRIG:SEQ3 TRIG:ACQ |
Sequence 3 group software trigger |
SENS:APER | Sets the aperture time (measurement time) |
ABOR | Aborts the trigger function |
MEAS:CURR:DC FETC:CURR:DC |
Queries the DC current |
MEAS:CURR:AC FETC:CURR:AC |
Queries the AC current |
MEAS:CURR:AMP:MAX FETC:CURR:AMP:MAX |
Queries the peak current |
SENS:CURR:HOLD | Sets the hold time of the peak current |
MEAS:CURR:AMP:MAX:HOLD FETC:CURR:AMP:MAX:HOLD |
Queries the peak current (the held value) |
SENS:CURR:PEAK:CLE | Clears the held measured peak current |
MEAS:CURR:CRES FETC:CURR:CRES |
Queries the current crest factor |
MEAS:CURR:HARM FETC:CURR:HARM |
Queries the harmonic current |
MEAS:CURR:HARM:RAT FETC:CURR:HARM:RAT |
Queries the harmonic current (as a percentage) |
MEAS:FREQ FETC:FREQ |
Queries the frequency |
MEAS:POW:AC FETC:POW:AC |
Queries the AC power (the actual power) |
MEAS:POW:AC:APP FETC:POW:AC:APP |
Queries the AC power (the apparent power) |
MEAS:POW:AC:REAC FETC:POW:AC:REAC |
Queries the AC power (the reactive power) |
MEAS:POW:AC:PFAC FETC:POW:AC:PFAC |
Queries the power factor of the AC power |
MEAS:POW:DC FETC:POW:DC |
Queries the DC power |
MEAS:VOLT:DC FETC:VOLT:DC |
Queries the DC voltage |
MEAS:VOLT:AC FETC:VOLT:AC |
Queries the AC voltage |
MEAS:VOLT:AMPL:MAX FETC:VOLT:AMPL:MAX |
Queries the voltage peak |
Measured Value Queries
(Single-phase, three-wire output/ Three-phase output)
MEAS{1|2}:CURR:DC FETC{1|2}:CURR:DC |
Queries the DC current of U/V phase (2P05-PCR-LE only) |
---|---|
MEAS{1|2|3}:CURR:AC FETC{1|2|3}:CURR:AC |
Queries the AC current of each phase |
MEAS{1|2|3}:CURR:AMPL:MAX FETC{1|2|3}:CURR:AMPL:MAX |
Queries the peak current of each phase |
MEAS{1|2|3}:CURR:AMP:MAX:HOLD FETC{1|2|3}:CURR:AMP:MAX:HOLD |
Queries the peak current (the held value) of each phase |
MEAS{1|2|3}:CURR:CRES FETC{1|2|3}:CURR:CRES |
Queries the current crest factor of each phase |
MEAS{1|2|3}:CURR:HARM FETC{1|2|3}:CURR:HARM |
Queries the harmonic current of each phase |
MEAS{1|2|3}:CURR:HARM:RAT FETC{1|2|3}:CURR:HARM:RAT |
Queries the harmonic current (as a percentage) of each phase |
MEAS{1|2}:LVOL:DC FETC{1|2}:LVOL:DC |
Queries the DC line voltage of each phase (2P05-PCR-LE only) |
MEAS{1|2|3}:LVOL:AC FETC{1|2|3}:LVOL:AC |
Queries the AC line voltage of each phase |
MEAS{1|2|3}:LVOL:AMPL:MAX FETC{1|2|3}:LVOL:AMPL:MAX |
Queries the peak AC line voltage of each phase |
MEAS{1|2|3}:POW:AC FETC{1|2|3}:POW:AC |
Queries the AC power (the actual power) of each phase |
MEAS:POW:AC:TOT FETC:POW:AC:TOT |
Queries total of the AC power (the actual power) |
MEAS{1|2|3}:POW:AC:APP FETC{1|2|3}:POW:AC:APP |
Queries the AC power (the apparent power) of each phase |
MEAS:POW:AC:APP:TOT FETC:POW:AC:APP:TOT |
Queries total of the AC power (the apparent power) |
MEAS{1|2|3}:POW:AC:REAC FETC{1|2|3}:POW:AC:REAC |
Queries the AC power (the reactive power) |
MEAS:POW:AC:REAC:TOT FETC:POW:AC:REAC:TOT |
Queries total of the AC power (the overall reactive power) |
MEAS{1|2|3}:POW:AC:PFAC FETC{1|2|3}:POW:AC:PFAC |
Queries the power factor of the AC power of each phase |
MEAS:POW:AC:PFAC:TOT FETC:POW:AC:PFAC:TOT |
Queries total of the power factor of the AC power |
MEAS{1|2}:POW:DC FETC{1|2}:POW:DC |
Queries the DC power of U/V phase (2P05-PCR-LE only) |
MEAS:POW:DC:TOT FETC:POW:DC:TOT |
Queries total of the DC power (2P05-PCR-LE only) |
MEAS{1|2}:VOLT:DC FETC{1|2}:VOLT:DC |
Queries the DC voltage of U/V phase (2P05-PCR-LE only) |
MEAS{1|2|3}:VOLT:AC FETC{1|2|3}:VOLT:AC |
Queries the AC voltage of each phase |
MEAS{1|2|3}:VOLT:AMPL:MAX FETC{1|2|3}:VOLT:AMPL:MAX |
Queries the voltage peak of each phase |