PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

Overview of commands

Default State

When the PLZ12005WH/PLZ20005WH is switched to the following default states, there are settings whose values are changed.

  • *RST command transmission
  • *RCL command transmission
  • Power on
  • Factory default (SYST:SEC:IMM transmission)

The following tables list the values of the settings that change after they have changed.

SOURce subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
ARB:COUN 3 Depends on the settings in memory Depends on PON setting 3 --
COND 0.0 Depends on the settings in memory Depends on PON setting 0.0 SIE
COND:EXT:FCON OFF Depends on the settings in memory Depends on PON setting OFF --
COND:PULS:LEV 0.0 Depends on the settings in memory Depends on PON setting 0.0 SIE
COND:PULS:FREQ 1.0 Depends on the settings in memory Depends on PON setting 1.0 HZ
COND:PULS:DCYC 50 Depends on the settings in memory Depends on PON setting 50 PCT
COND:RANG HIGH Depends on the settings in memory Depends on PON setting HIGH --
COND:RESP NORM Depends on the settings in memory Depends on PON setting NORM --
CURR 0.0 Depends on the settings in memory Depends on PON setting 0.0 A
CURR:EXT:FCON OFF Depends on the settings in memory Depends on PON setting OFF --
CURR:EXT:ACON OFF Depends on the settings in memory Depends on PON setting OFF --
CURR:PULS:LEV 0.0 Depends on the settings in memory Depends on PON setting 0.0 A
CURR:PULS:DCYC 50 Depends on the settings in memory Depends on PON setting 50 PCT
CURR:PULS:FREQ 1.0 Depends on the settings in memory Depends on PON setting 1.0 Hz
CURR:PROT MAX Depends on the settings in memory Depends on PON setting MAX A
CURR:PROT:STAT OFF Depends on the settings in memory Depends on PON setting OFF --
CURR:SINE:AMPL 0.0 Depends on the settings in memory Depends on PON setting 0.0 A
CURR:SINE:FREQ 1.0 Depends on the settings in memory Depends on PON setting 1.0 HZ
CURR:SLEW MIN Depends on the settings in memory Depends on PON setting MIN A/μs
CURR:SST MIN Depends on the settings in memory Depends on PON setting MIN S
FUNC CC Depends on the settings in memory Depends on PON setting CC --
FUNC:CVOP OFF Depends on the settings in memory Depends on PON setting OFF --
POW 0.0 Depends on the settings in memory Depends on PON setting 0.0 V
POW:EXT:FCON OFF Depends on the settings in memory Depends on PON setting OFF --
POW:PROT MAX Depends on the settings in memory Depends on PON setting MAX W
POW:PROT:STAT OFF Depends on the settings in memory Depends on PON setting OFF --
PROG ""(No selection) --
VOLT 0.0 Depends on the settings in memory Depends on PON setting 0.0 V
VOLT:EXT:FCON OFF Depends on the settings in memory Depends on PON setting OFF --
VOLT:PROT:ACT LIM Depends on the settings in memory Depends on PON setting LIM --
VOLT:PROT:LOW 0.0 Depends on the settings in memory Depends on PON setting 0.0 V
VOLT:PROT:STAT OFF Depends on the settings in memory Depends on PON setting OFF --
VOLT:RESP NORM Depends on the settings in memory Depends on PON setting NORM --

OUTPut/INPut subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
INP
OUTP
OFF Depends on the settings in memory Depends on PON setting OFF --
INP:CUTO:CAP
OUTP:CUTO:CAP
0 Depends on the settings in memory Depends on PON setting 0 A
INP:CUTO:CAP:STAT
OUTP:CUTO:CAP:STAT
OFF Depends on the settings in memory Depends on PON setting OFF --
INP:CUTO:ENER
OUTP:CUTO:ENER
0 Depends on the settings in memory Depends on PON setting 0 A
INP:CUTO:ENER:STAT
OUTP:CUTO:ENER:STAT
OFF Depends on the settings in memory Depends on PON setting OFF A
INP:CUTO:ETIM
OUTP:CUTO:ETIM
0 Depends on the settings in memory Depends on PON setting 0 A
INP:CUTO:ETIM:STAT
OUTP:CUTO:ETIM:STAT
OFF Depends on the settings in memory Depends on PON setting OFF --
INP:CUTO:VOLT
OUTP:CUTO:VOLT
0 Depends on the settings in memory Depends on PON setting 0 A
INP:CUTO:VOLT:STAT
OUTP:CUTO:VOLT:STAT
OFF Depends on the settings in memory Depends on PON setting OFF --
INP:EXT:LOG
OUTP:EXT:LOG
POS Depends on the settings in memory Depends on PON setting POS --
INP:MSYN:ACC
OUTP:MSYN:ACC
OFF Depends on the settings in memory Depends on PON setting OFF --
INP:PON:STAT
OUTP:PON:STAT
No change Value at previous POWER off AUTO --
INP:PROT:WDOG
OUTP:PROT:WDOG
No change Value at previous POWER off OFF --
INP:PROT:WDOG:DEL
OUTP:PROT:WDOG:DEL
No change Value at previous POWER off 60 S

TRIG subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
TRIG:ACQ:COUN 1 Depends on the settings in memory Depends on PON setting 1 --
TRIG:ACQ:DEL 0 Depends on the settings in memory Depends on PON setting 0 S
TRIG:ACQ:SOUR IMM Depends on the settings in memory Depends on PON setting IMM --
TRIG:ACQ:INT:STAT OFF Depends on the settings in memory Depends on PON setting OFF --
TRIG:ACQ:INT:TIME 0.1 Depends on the settings in memory Depends on PON setting 0.1 S
TRIG:TRAN:DEL 0.0 Depends on the settings in memory Depends on PON setting 0.0 S
TRIG:TRAN:SOUR IMM Depends on the settings in memory Depends on PON setting IMM --

DATA subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
DATA:BSIZ 1024 Depends on the settings in memory Depends on PON setting 1024 --
DATA:FORM CURR,VOL,POW Depends on the settings in memory Depends on PON setting CURR,VOL,POW --
DATA:INT:GATE LOAD ON Depends on the settings in memory Depends on PON setting LOAD ON --
DATA:INT:GATE:ARES ON Depends on the settings in memory Depends on PON setting ON --

SYSTem subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
SYST:BEEP No change Value at previous POWER off ON --
SYST:BEEP:KEY No change Value at previous POWER off ON --
SYST:BEEP:PROT No change Value at previous POWER off ON --
SYST:BEEP:SCPI No change Value at previous POWER off ON --
SYST:COMM:RLST No change LOCal --
SYST:CONF:DIGITAL2:DIR No change Value at previous POWER off INP --
SYST:CONF:RSEN No change Value at previous POWER off OFF --
SYST:KLOC No change Value at previous POWER off OFF --
SYST:KLOC:LEV No change Value at previous POWER off 3 --
SYST:PASS:NEW No change Value at previous POWER off "" --
SYST:SSAV No change Value at previous POWER off OFF --
SYST:SSAV:DEL No change Value at previous POWER off 60 S
SYST:TZON No change No change UTC --

STATus subsystem

Setting item Value Unit
*RST *RCL Power on Factory default
STAT:OPER:ENAB No change 0 --
STAT:OPER:PTR No change 32767 --
STAT:OPER:NTR No change 0 --
STAT:QUES:ENAB No change 0 --
STAT:QUES:PTR No change 32767 --
STAT:QUES:NTR No change 0 --

Other subsystems

Setting item Value Unit
*RST *RCL Power on Factory default
DISP:ETIM OFF Depends on the settings in memory Depends on PON setting OFF --
DISP:CAP OFF Depends on the settings in memory Depends on PON setting OFF --
DISP:ENER OFF Depends on the settings in memory Depends on PON setting OFF --
SENS:APER 0.1 Depends on the settings in memory Depends on PON setting 0.1 S

IEEE 488.2 common command

Setting item Value Unit
*RST *RCL Power on Factory default
*ESE No change Depends on PSC setting 0 --
*PSC No change Value at previous POWER off ON --
*SRE No change Depends on PSC setting 0 --

>top

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ12005WH/PLZ20005WHElectronic Load PLZ12005WH/PLZ20005WH
Communication Interface Manual