PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

TRIG:TRAN:SOUR

Sets the trigger source for clearing sequence start trigger wait states.

Command

TRIGger:TRANsient:SOURce <character>

TRIGger:TRANsient:SOURce?

Parameter <character>

You can select from the available choices below.

Value IMMediate Starts the sequence when the Initiate key on the PLZ12005WH/PLZ20005WH front panel is pressed or when INIT:TRAN:PROG is received from a PC or the like.
BUS Starts execution at the timing of software trigger application by the 4882/SCPI command
DIGITAL2 Starts execution at the timing at which a signal is input to pin 13 of the EXT CONT connector of the PLZ12005WH/PLZ20005WH (when SYST:CONF:DIGITAL2:DIR is set to INPut).
MSYNc Starts the sequence when an MSync key on the front panel of the synchronized PLZ12005WH/PLZ20005WHs is pressed. Starts the sequence when TRIG:TRAN:MSYN is received from a PC or the like.
LOADOFF Executes the sequence when the load is turned off.
*RST transmission IMMediate  

Example

TRIG:TRAN:SOUR DIGITAL2

Response

Returns the source of the trigger to be applied for TRANsient in character format in response to the TRIG:TRAN:SOUR? query.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ12005WH/PLZ20005WHElectronic Load PLZ12005WH/PLZ20005WH
Communication Interface Manual