PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

CURR:SINE:FREQ

Sets the frequency of the sine function.

Command

[SOURce:]CURRent[:LEVel]:SINE:FREQuency {<numeric>|MINimum|MAXimum}

[SOURce:]CURRent[:LEVel]:SINE:FREQuency? [{MINimum|MAXimum}]

Parameter

Setting range 1 to 1k, 2k, 5k, 10k  
Resolution 1 to 10 1
20 to 100 10
200 to 1000 100
Unit Hz  
*RST transmission 1  

Response

Returns the frequency of the sine function in NR3 format in response to the CURR:SINE:FREQ? query.

 

(Reference) The period resolution ⊿T is set to 20 [μs] in the device. The response returns the inverse of the period set in the device. For example, if you specify 900 Hz, 1120 [μs] is set in the device. The response in this situation will be 892.86Hz.