PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

FUNC:CVOP

Enables/disables +CV mode (addition of CV mode to CC mode or CR mode).

+CV mode can only be set with an SCPI command. This command is available to maintain compatibility with the PLZ-5W series (PLZ205W/PLZ405W/PLZ1205W).


If +CV mode is enabled, the UVP action (PROGram:VOLTage:PROTection:ACTion) is set to LIMit. The LIMit value changes according to the changes in the voltage (VOLTage) in CV mode.


If +CV mode is disabled, the UVP action is set to OFF (disabled) regardless of TRIP/LIMit setting.


Command

[SOURce:]FUNCtion[:MODE]:CVOPtion[:STATe] {ON|OFF|1|0}

[SOURce:]FUNCtion[:MODE]:CVOPtion[:STATe]?

Parameter

Value ON(1) Enabled
OFF(0) Disabled
*RST transmission OFF(0)  

Response

Returns +CV mode enable/disable in boolean format in response to the FUNC:CVOP? query.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ12005WH/PLZ20005WHElectronic Load PLZ12005WH/PLZ20005WH
Communication Interface Manual