PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

CURR:PULS:FREQ

Sets the pulse frequency. (Alias for COND:PULS:FREQ)

This frequency is used in common for CURRent and CONDuctance pulse.

Command

[SOURce:]CURRent[:LEVel]:PULSe:FREQuency {<numeric>|MINimum|MAXimum}

[SOURce:]CURRent[:LEVel]:PULSe:FREQuency? [{MINimum|MAXimum}]

Parameter

Setting range 1.0 to 10.0k  
Resolution 1 to 10 0.1
11 to 100 1
110 to 1000 10
1.1k to 10.0k 0.1k
Unit Hz  
*RST transmission 1  

Response

Returns the pulse frequency in NR3 format in response to the CURR:PULS:FREQ? query.

 

(Reference) The period resolution ⊿T is set to 1 [μs] in the device. The response returns the inverse of the period set in the device. For example, if you specify 9300 Hz, 108 [μs] is set in the device. The response in this situation will be 9259.26Hz.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ12005WH/PLZ20005WHElectronic Load PLZ12005WH/PLZ20005WH
Communication Interface Manual