Command (function search)
Trigger Function
The PLZ12005WH/PLZ20005WH has 2 independent triggers (TRANsient, ACQuire). →Overview of Trigger Function
Trigger Function (ACQuire)
Starts measurement at trigger application.
INIT:ACQuire | Switches to measurement trigger wait state |
---|---|
TRIG:ACQ:COUN | Sets the number of times measurement values are to be recorded |
TRIG:ACQ:DEL | Sets the delay time for applying measurement triggers |
TRIG:ACQ:SOUR | Sets the measurement trigger source |
TRIG:ACQ:TNT:STAT | Sets whether to measure at intervals |
TRIG:ACQ:TNT:TIME | Sets the measurement interval time |
TRIG:ACQ:MSYN | Starts measurement simultaneously on synchronized PLZ12005WH/PLZ20005WHs |
ABOR:ACQ | Stops the measurement trigger function |
Trigger Function (TRANsient)
Starts sequence at trigger application.
INIT:TRAN:PROG | Switches to the sequence trigger wait state in the program selected by PROG |
---|---|
TRIG:TRAN:DEL | Sets the delay time for applying sequence triggers |
TRIG:TRAN:SUSP | Pauses a sequence that is being executed |
TRIG:TRAN:RES | Resumes a sequence that has been paused |
TRIG:TRAN:EXEC? | Queries the execution state of the sequence trigger function and the pulse function |
TRIG:TRAN:SOUR | Sets the sequence start trigger source |
TRIG:TRAN:MSYN | Starts a sequence simultaneously on synchronized PLZ12005WH/PLZ20005WHs |
ABOR:TRAN | Stops the sequence trigger function |
Trigger functions (common to ACQuire, TRANsient)
*TRG | Sends a software trigger |
---|---|
ABOR | Stops all the trigger functions |
Related information