PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

TRIG:ACQ:SOUR

Sets the trigger source for clearing sequence start trigger wait states.

Command

TRIGger:ACQuire:SOURce <character>

TRIGger:ACQuire:SOURce?

Parameter <character>

You can select from the available choices below.

Value IMMediate Executes the measurement values when the Initiate key on the PLZ12005WH/PLZ20005WH front panel is pressed or when INIT:ACQ is received from a PC or the like.
BUS Records the measurement values at the timing of software trigger (*TRG) application. The trigger is applied to both trigger systems: (1) sequence and step (TRANsient) and (2) measure (ACQuire).
DIGITAL2 Executes the measurement values at the timing at which a signal is input to pin 13 of the EXT CONT connector of the PLZ12005WH/PLZ20005WH (when SYST:CONF:DIGITAL2:DIR is set to INPut).
TALink If PROG:STEP<n>:TRIG:GEN is set to TALink, measured values are recorded during the step execution of the sequence.
MSYNc Executes the step when TRIG:ACQ:MSYN is received from a PC or the like.
LOADOFF Executes the sequence when the load is turned off.
*RST transmission IMMediate  

Example

TRIG:ACQ:SOUR DIGITAL2

Response

Returns the source of the trigger of the ACQuire subsystem in character format in response to the TRIG:ACQ:SOUR? query.