PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

CURR:PULS:DCYC

Sets the pulse duty cycle.

This frequency is used in common for CURRent and CONDuctance pulse.

Command

[SOURce:]CURRent[:LEVel]:PULSe:DCYCle {<numeric>|MINimum|MAXimum}

[SOURce:]CURRent[:LEVel]:PULSe:DCYCle? [{MINimum|MAXimum}]

Parameter

Duty cycle setting
range, step
1Hz to 10Hz 5.0 to 95.0, 0.1 step
11Hz to 100Hz
110Hz to 1000Hz
1.1kHz to 10.0kHz 5 to 95, 1 step
Unit %  
*RST transmission 50  

Response

Returns the pulse duty cycle in NR3 format in response to the CURR:PULS:DCYC? query.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ12005WH/PLZ20005WHElectronic Load PLZ12005WH/PLZ20005WH
Communication Interface Manual