Command (function search)
Synchronized Operation
Synchronized operation is possible among multiple synchronized PLA-5Ws.
For details on how to establish synchronized connections, see the user’s manual contained in the accompanying CD-ROM.
Synchronizing load on/off
INP:MSYN:ACC | Sets whether to accept load on/off sync signal (IMP:MSYN) input from other synchronized devices |
---|---|
INP:MSYN | Turns the load on or off simultaneously on synchronized PLZ12005WH/PLZ20005WHs |
Synchronizing measurements
TRIG:ACQ:SOUR | Measurement trigger source. To synchronize, specify MSYNc. |
---|---|
INIT:ACQ | Switches to measurement trigger wait state |
TRIG:ACQ:MSYN | Starts measurement simultaneously on synchronized PLZ12005WH/PLZ20005WHs |
Synchronizing the start of sequences
TRIG:TRAN:SOUR | Sequence trigger source. To synchronize, specify MSYNc. |
---|---|
INIT:TRAN:PROG | Switches to the sequence trigger wait state in the program selected by PROG |
TRIG:TRAN:MSYN | Starts a sequence simultaneously on synchronized PLZ12005WH/PLZ20005WHs |
Synchronizing the resuming of sequences
PROG:STEP<n>:TRIG:WAIT | Sets trigger wait at the specified step of the program selected by PROG and sets the trigger source To synchronize, set the trigger source to MSYNc. |
---|---|