AC Power Supply PCR-LE series

AC Power Supply PCR-LE2 series

AC Power Supply PCR-M series

Commnunication Interface Mannual


 

SYST:CONF:TPH:MODE

Sets the output mode for single-phase, three-phase output.

This command is valid in AC mode (OUTP:COUP AC).

This command is not valid when the output is on (OUTP ON).

This command is valid in single-phase three-wire output (PCR-LE optional).

Command

SYSTem:CONFigure:TPHase:MODE {ON|OFF|1|0}

SYSTem:CONFigure:TPHase:MODE?

Parameter

Value ON(1) Double-phase output
  OFF(0) Single-phase three-wire output

For the setting that is applied when *RST is sent, see this table

For the setting that is applied when *RST is sent, see this table

For the setting that is applied when *RST is sent, see this table

Response

Response

Returns the output mode in NR1 format in response to SYST:CONF:TPH:MODE?.