AC Power Supply PCR-LE series

AC Power Supply PCR-LE2 series

AC Power Supply PCR-M series

Commnunication Interface Mannual


 

MEM:RCL

The following settings are recalled from memory.

  AC mode DC mode AC+DC mode
Frequency*1 Yes No Yes
AC voltage*2 Yes No Yes*3
DC voltage*2 No Yes*4 Yes*3
Wave bank No. Yes No Yes

*1: If the frequency that you recall is lower than the frequency lower limit, the frequency is set to the lower limit.
If the frequency that you recall is higher than the frequency upper limit, the frequency is set to the upper limit.

*2: If the voltage that you recall is lower than the voltage lower limit or is higher than the voltage upper limit, the voltage is set to 0 V.
If the L range is in use and a voltage that is only valid in the H range is recalled, the voltage is set to 0 V.

*3: If the peak waveform value is outside of -431 V to 431 V for the H range or -215.5 V to 215.5 V for the L range, the voltage is set to 0 V.

*4: During single-phase, three-wire output (PCR-LE optional), the V phase is automatically set to the same amplitude as the U phase but with opposite polarity.

You can view the settings that are stored in memory by using the MEM:PREV command.

If the EX05-PCR-LE is in use (optional) and the signal source is set to external signal source (FUNC:SOUR EXT), this command is invalid.

Command

MEMory:RCL <NR1>

Parameter

Value 0 to 99 Memory No.

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

AC power supply  PCR-M seriesAC Power Supply PCR-LE series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-LE2 series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-M series
Communication Interface Manual