MEAS:POW:AC:TOT/ FETC:POW:AC:TOT
MEAS:POW:AC:TOT? starts a new measurement and then queries the total AC power.
FETC:POW:AC:TOT? queries the total AC power without starting a new measurement.
This command is valid in single-phase three-wire output or three-phase output (PCR-LE optional).
The valid modes (parameters to the OUTP:COUP command) is AC.
Command
MEASure[:SCALar]:POWer:AC:TOTal?
FETCh[:SCALar]:POWer:AC:TOTal?
Response
In case of the single-phase three-wire output, returns the total AC power of U and V phases in NR3 format in response to MEAS:POW:AC:TOT? of FETC:POW:AC:TOT?.
In case of the three-phase output, returns the total AC power of U, V, and W phases in NR3 format in response to MEAS:POW:AC:TOT? of FETC:POW:AC:TOT?.
Unit: W