MEAS:POW:AC:REAC:TOT/ FETC:POW:AC:REAC:TOT
MEAS:POW:AC:REAC:TOT? starts a new measurement and then queries the AC power (the overall reactive power).
FETC:POW:AC:REAC:TOT? queries the AC power (the overall reactive power) without starting a new measurement.
The overall reactive power is calculated from the total of the apparent power of each phase (Stotal) and the total of the active power of each phase (Ptotal).
This command is valid in single-phase three-wire output or three-phase output (PCR-LE optional).
The valid modes (parameters to the OUTP:COUP command) are AC.
Command
MEASure[0][:SCALar]:POWer:AC:REACtive:TOTal?
FETCh[0][:SCALar]:POWer:AC:REACtive:TOTal?
Response
Returns the AC power (the overall reactive power) in NR3 format in response to MEAS:POW:AC:REAC:TOT? or FETC:POW:AC:REAC:TOT?.
Unit:VAR