AC Power Supply PCR-LE series

AC Power Supply PCR-LE2 series

AC Power Supply PCR-M series

Commnunication Interface Mannual


 

MEAS{1|2|3}:CURR:HARM/ FETC{1|2|3}:CURR:HARM

MEAS{1|2|3}:CURR:HARM? starts a new measurement and then queries the harmonic currents of the specified phase.

FETC{1|2|3}:CURR:HARM? queries the harmonic currents of the specified phase without starting a new measurement.

This command is valid in single-phase three-wire output or three-phase output (PCR-LE optional).

If the EX05-PCR-LE is in use (optional), this command is valid when the signal source is set to internal signal source (FUNC:SOUR INT).

The valid modes (parameters to the OUTP:COUP command) is AC.

Command

U phase

MEASure1[:SCALar]:CURRent:HARMonic?

FETCh1[:SCALar]:CURRent:HARMonic?

V phase

MEASure2[:SCALar]:CURRent:HARMonic?

FETCh2[:SCALar]:CURRent:HARMonic?

W phase (Three-phase output only)

MEASure3[:SCALar]:CURRent:HARMonic?

FETCh3[:SCALar]:CURRent:HARMonic?

Response

Returns the harmonic currents of U phase, from the first to the fortieth, in a comma-separated list of values in NR3 format in response to MEAS1:CURR:HARM? or FETC1:CURR:HARM?.

Returns the harmonic currents of V phase, from the first to the fortieth, in a comma-separated list of values in NR3 format in response to MEAS2:CURR:HARM? or FETC2:CURR:HARM?.

Returns the harmonic currents of W phase, from the first to the fortieth, in a comma-separated list of values in NR3 format in response to MEAS3:CURR:HARM? or FETC3:CURR:HARM?.

Unit: A

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

AC power supply  PCR-M seriesAC Power Supply PCR-LE series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-LE2 series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-M series
Communication Interface Manual