INIT:NAME
Starts the trigger function of the specified sequence group.
If the trigger source is set to IMM, the settings are changed immediately, or the measurement, power line abnormality simulation, or sequence starts immediately. If the trigger source is set to BUS, the PCR waits for a software trigger, and then the settings are changed, or the measurement, power line abnormality simulation, or sequence starts.
Command
INITiate[:IMMediate]:NAME {TRANsient|ACQuire|SIMulation|PROGram}
Parameter
Value | TRAN | Starts the output change (sequence 1) trigger function |
ACQ | Starts the measurement (sequence 3) trigger function | |
SIM | Starts the power line abnormality simulation trigger function (sequence 4; PCR-LE only) This is not valid when the action that is performed when the current limit is exceeded is set to CC (CURR:PROT:STAT OFF). |
|
PROG | Starts the sequence operation trigger function (sequence 5; PCR-LE only) This is not valid when the action that is performed when the current limit is exceeded is set to CC (CURR:PROT:STAT OFF). |