PROG:EDIT (PCR-LE/ PCR-LE2 only)
Configures a sequence step (excluding the jump and impedance settings).
Command
PROGram:EDIT <step_NR1>,<freq-ramp_bool>,<freq_numeric>,
<acv-ramp_bool>,<acv_numeric>,<dcv-ramp_bool>,<dcv_numeric>,<time_numeric>,<bank_NR1>,
<status-out_bool>,<trig-out_bool>,<trig-in_bool>,<output_bool>
PROGram:EDIT? <step_NR1>
Parameter <step_NR1>
Value | 0 to 599 | Number of the step that you want to configure |
Parameter <freq-ramp> {ON|OFF|1|0}
Value | ON(1) | Enables the ramped frequency transition |
OFF(0) | Disables the ramped frequency transition |
Parameter <freq> {<numeric>|MAXimum|MINimum}
Value | 1 to 999.9 | Frequency |
Unit | HZ |
Parameter <acv-ramp> {ON|OFF|1|0}
Value | ON(1) | Enables the ramped AC voltage transition |
OFF(0) | Disables the ramped AC voltage transition |
Parameter <acv> {<numeric>|MAXimum|MINimum}
Value | 0 to 305.0 | AC voltage*1 |
Unit | V |
*1 During single-phase, three-wire output and three-phase output (optional), all phases are set at the same time.
Parameter <dcv-ramp> {ON|OFF|1|0}
Value | ON(1) | Enables the ramped DC voltage signal change |
OFF(0) | Disables the ramped DC voltage signal change |
Parameter <dcv> {<numeric>|MAXimum|MINimum}
Value | 0 to ±431.0 | DC voltage (About the values for single-phase, three-wire output and three-phase output) |
Unit | V |
Parameter <time> {<numeric>|MAXimum|MINimum}
Value | 0.0001s to 1000h | Step time |
Unit | S |
Parameter <bank_NR1>
Value | 0 to 63 | Number of the waveform bank to use |
Parameter <status-out> {ON|OFF|1|0}
Value | ON(1) | Enables status output |
OFF(0) | Disables status output |
Parameter <trig-out> {ON|OFF|1|0}
Value | ON(1) | Enables trigger output |
OFF(0) | Disables trigger output |
Parameter <trig-in> {ON|OFF|1|0}
Value | ON(1) | Enables trigger input |
OFF(0) | Disables trigger input |
Parameter <output> {ON|OFF|1|0}
Value | ON(1) | Enables output |
OFF(0) | Disables output |
Response
Returns the settings of the specified step in the following order in response to PROG:EDIT? <step_NR1>.
Frequency signal change <NR1>,frequency <NR3>,AC voltage signal change <NR1>,AC voltage*2 <NR3>, DC voltage signal change <NR1>, DC voltage*3 <NC3>, step execution time <NR3>, waveform bank number <NR1>, status output <NR1>, trigger output <NR1>, trigger input <NR1>, output on/off <NR1>
*2 During single-phase, three-wire output and three-phase output (optional), the U phase AC voltage is returned when the unbalanced voltage is set.
*3 During two-phase output and three-phase output (optional), the U phase DC voltage is returned when the unbalanced voltage is set.