AC Power Supply PCR-LE series

AC Power Supply PCR-LE2 series

AC Power Supply PCR-M series

Commnunication Interface Mannual


 

OPERation status register (PCR-M)

The OPERation status register is a 16-bit register that contains conditions that are part of the PCR normal operations.

 

Bit Bit weight Bit name Description
0 1 NOT USED --
1 2 NOT USED --
2 4 NOT USED --
3 8 NOT USED --
4 16 MEASuring *1 Indicates whether measurement is in progress on the PCR.
5 32 Waiting for TRIGger *2 Indicates whether the PCR is waiting for a trigger (TRIG)
6 64 NOT USED --
7 128 NOT USED --
8 256 CV CV output
9 512 NOT USED --
10 1024 NOT USED --
11 2048 NOT USED --
12 4096 NOT USED --
13 8192 NOT USED --
14 16384 NOT USED --
15 32768 NOT USED Always zero

*1. A summary bit of the STAT:OPER:MEAS register. Further detailed information can be obtained by referring to this register.

*2. A summary bit of the STAT:OPER:TRIG register. Further detailed information can be obtained by referring to this register.

>top

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

AC power supply  PCR-M seriesAC Power Supply PCR-LE series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-LE2 series
Communication Interface Manual

AC power supply  PCR-M seriesAC Power Supply PCR-M series
Communication Interface Manual