Interface Setup
Interface Setup for PCR-M series
USB (optional)
This interface is valid only when the optional USB interface board is installed.
A device driver supporting USB T&M Class (USBTMC) is required to control the PCR through the USB interface. The USBTMC driver is automatically installed by the VISA library.
USB connection
Use a standard USB cable to connect the PCR to the computer.
USB configuration
1. Press CONFIG until the interface is set to "IntF."
2. Turn the rotary knob to set USB to "USb."
3. Wait at least 5 seconds, and then turn the POWER switch off and on.
The settings are applied.
Service request
The PCR is equipped with service request and serial polling functions.
USB function
Complies with USB Specification 2.0
Complies with USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0
Data rate: 12 Mbps maximum (full speed)
VID (Vender ID)
0x0B3E
PID (Product ID)
PCR500M: 0x1009
PCR1000M: 0x100A
PCR2000M: 0x100B
PCR4000M 0x1043