Interface Setup
Interface Setup for PCR-LEPCR-LE2 series
USB(optional)
This interface is valid only when the optional USB interface board is installed.
Attachment of the interface board
A device driver supporting USB T&M Class (USBTMC) is required to control the PCR through the USB interface. The USBTMC driver is automatically installed by the VISA library.
USB connection
Use a standard USB cable to connect the PCR to the computer.
USB configuration
1. Press CONFIG (SHIFT+OPR MODE), COM-I/F, TYPE, and then USB.
2. Restart the PCR-LEPCR-LE2.
The settings are applied.
Service request
The PCR is equipped with service request and serial polling functions.
USB function
Complies with USB Specification 2.0
Complies with USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0
Data rate: 12 Mbps maximum (full speed)
VID (Vender ID)
0x0B3E
PID (Product ID)
PCR6000LE2: 0x103A
PCR9000LE2: 0x103B
PCR12000LE2: 0x103C
PCR18000LE2: 0x103E
PCR27000LE2: 0x1041