PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

PROG:STEP<n>:TRIG:GEN

Sets whether to perform trigger output at the specified step of the program selected by PROG.

Command

[SOURce:]PROGram[:SELected]:STEP<n>:TRIGger:GENerate <character>[,<character>,<character>,<character>,<character>]

[SOURce:]PROGram[:SELected]:STEP<n>:TRIGger:GENerate?

Parameter <n>

Value Specifies the step number
Example STEP1

Parameter <character>

If NONE is set, it cannot be specified at the same time as other setting values.
For all other setting values, up to 5 items can be set.

Value NONE Trigger not generated
TRIGOUT Trigger signal output to TRIGOUT pin
DIGITAL0 Trigger signal output to DIGITAL0 pin
DIGITAL1 Trigger signal output to DIGITAL1 pin
DIGITAL2 Trigger signal output to DIGITAL2 pin (When I/O is OUTPut)
TALink Output TALink triggers. If TRIG:ACQ:SOUR is set to TALink, measured values are recorded during the step execution of the sequence.

Example

PROG:STEP1:TRIG:GEN TRIGOUT,DIGITAL2,TAL

Response

Returns the trigger output setting at the specified step in character format in response to the PROG:STEP<n>:TRIG:GEN? query.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)