PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

PROG:CURR:PROT:ACT

Sets the operation to be performed when overcurrent protection (OCP) is activated in the program selected by PROG. (Alias for PROG:CURR:PROT:STAT)

Command

[SOURce:]PROGram[:SELected]:CURRent:PROTection:ACTion {TRIP|LIMit}

[SOURce:]PROGram[:SELected]:CURRent:PROTection:ACTion?

Parameter

Value TRIP Load off
LIMit Limit output
Default value LIMit  

Response

Returns the action to be performed when OCP in the selected program is activated in character format in response to the PROG:CURR:PROT:ACT? query.