PLZ12005WH/PLZ20005WH Electronic Load

Commnunication Interface Mannual


 

Setup

Setting Up the LAN Interface

WARNING

If a network problem occurs, an unexpected dangerous voltage may occur that may cause electric shock, fire, physical damage to the DUT, and so on. If you are going to remotely control the PLZ12005WH/PLZ20005WH from a distance, install a Web camera or take other measures to monitor the status.

Either the VXI-11, HiSLIP, or SCPI-RAW protocol can be used to control the PLZ12005WH/PLZ20005WH via the LAN interface. For VXI-11 and HiSLIP, the middleware is installed automatically by the VISA library. For SCPI-RAW, no middleware is required.

The LAN interface board has a Web browser interface. You can configure the LAN interface settings from your PC's Web browser.

For information on topics such as connecting to your corporate LAN, your IP address, your host name, and security, contact your network administrator.

If you are using a host name (a Bonjour host name), you have to install Apple Bonjour.

Connection Procedure

Use a standard LAN cable (category 5 and straight) to connect the PLZ12005WH/PLZ20005WH to a network hub or router.

Setting Procedure

1.    Press SYSTEM and then Configure Interface on the front panel.

If the Interface screen does not appear, press Configure Interface again.

2.     Press Modify and make the necessary LAN settings for “IP Address”, “DNS Server”, “WINS Server”, and “Hostname & Services”.

Normally, use the PLZ12005WH/PLZ20005WH with “Method” under “IP Address” set to “Automatic” (factory default setting). If this settings have been changed, execute Resetting the LAN Interface Settings to restore the factory default settings.

You can check the IP address by pressing SYSTEM and then Information on the front panel.

WARNING

Possible damage to the equipment and electric shock. The LAN interface can be accessed from anywhere on the network that the PLZ12005WH/PLZ20005WH is connected to. If necessary, set a password for the Web browser interface.
->Setting Procedure

NOTE

If you are going to use remote control, we recommend that you set the PLZ12005WH/PLZ20005WH in remote mode (set SYST:COMM:RLST to REMote or RWLock). When operating in remote mode, REMOTE appears on the front panel display.

Service request

The PLZ12005WH/PLZ20005WH is equipped with service request and serial polling functions.

LAN feature

Depending on the operations that you perform through the Web browser interface, the PLZ12005WH/PLZ20005WH may need to connect to the Internet.

Complies with LXI 1.4 Core Specification 2011

Complies with the VXI-11, HiSLIP, SCPI-RAW, and SCPI-Telnet protocols

Baud rate: 100 Mbps maximum (auto negotiation)

AUTO MDIX support

From your Web browser, you can:

・Operating environment

Internet Explorer 8.0 or later, Mozilla Firefox 3.6 or later, Google Chrome 5.0 or later, Safari 5.1 or later,

・Functions that can be used

Measuring instrument information, network information, display of VISA resource information, checking the connected PLZ12005WH/PLZ20005WH, remote control from browser, changing network settings, system status information, license information, password setting

Resetting the LAN interface

When the LAN interface is reset, the Interface items other than “Desired Hostname”, “Desired Description”, “NTP Server Hostname”, and “Auto Clock Adjustment” are restored to the factory default settings. This operation does not affect the other settings.

You can use this if you forget the security password or IP address.

1.    Press SYSTEM and then Configure Interface on the front panel.

If the Interface screen does not appear, press Configure Interface again.

2.    Press LAN Reset and then ENTER.

The LAN interface setting is reset. If the PLZ12005WH/PLZ20005WH is in remote mode, reset will change the PLZ12005WH/PLZ20005WH back into local mode (panel operation).