VOLT:PROT:UND

Set the overvoltage protection.

Command

[SOURce:]VOLTage:PROTection[:LEVel]UNDer {<value>|MINimum|MAXimum}

[SOURce:]VOLTage:PROTection[:LEVel]UNDer?

Parameter (PLZ-4W)

Value LOW range 0.000 to 15.000 (Resolution: 0.001)
HIGH range 0.00 to 150.00 (Resolution: 0.01)
Default 0
Unit V

Table shows the setting that is specified when *RST or *RCL is sent.

If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.

Parameter (PLZ-4WL)

Value LOW range 0.3 to 4.0 (Resolution: 0.001)
HIGH range 0.3 to 30.0 (Resolution: 0.01)
Default 0.3
Unit V

Table shows the setting that is specified when *RST or *RCL is sent.

If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.

Parameter (PLZ-4WH)

Value LOW range 5.00 to 65.00 (Resolution: 0.01)
HIGH range 5.0 to 650.0 (Resolution: 0.1)
Default 5
Unit V

Table shows the setting that is specified when *RST or *RCL is sent.

If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.

Response

Returns the overvoltage protection setting in the NR3 form in response to VOLT:PROT:UND?