VOLT:PROT:UND
Set the overvoltage protection.
Command
[SOURce:]VOLTage:PROTection[:LEVel]UNDer {<value>|MINimum|MAXimum}
[SOURce:]VOLTage:PROTection[:LEVel]UNDer?
Parameter (PLZ-4W)
Value | LOW range | 0.000 to 15.000 (Resolution: 0.001) |
HIGH range | 0.00 to 150.00 (Resolution: 0.01) | |
Default | 0 | |
Unit | V |
Table shows the setting that is specified when *RST or *RCL is sent.
If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.
Parameter (PLZ-4WL)
Value | LOW range | 0.3 to 4.0 (Resolution: 0.001) |
HIGH range | 0.3 to 30.0 (Resolution: 0.01) | |
Default | 0.3 | |
Unit | V |
Table shows the setting that is specified when *RST or *RCL is sent.
If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.
Parameter (PLZ-4WH)
Value | LOW range | 5.00 to 65.00 (Resolution: 0.01) |
HIGH range | 5.0 to 650.0 (Resolution: 0.1) | |
Default | 5 | |
Unit | V |
Table shows the setting that is specified when *RST or *RCL is sent.
If a valid VOLT:PROT:STAT is specified when undervoltage protection is off, the VOLT:PROT:UND automatically sets MIN.
Response
Returns the overvoltage protection setting in the NR3 form in response to VOLT:PROT:UND?