CURR:PROT:ACT

Sets whether to turn off the load or limit the current when the OCP trips.

Command

[SOURce:]CURRent:PROTection:ACTion {LIMit|TRIP}

[SOURce:]CURRent:PROTection:ACTion?

Parameter

Value LIMit Limit the current when the OCP trips.
TRIP Turn off the load when the OCP trips.

Table shows the setting that is specified when *RST or *RCL is sent.

Response

Returns the OCP operation setting in the CHAR form in response to CURR:PROT:ACT?