CURR:PROT

Set the overcurrent protection.

Command

[SOURce:]CURRent:PROTection[:LEVel][:OVER] {<value>|MINimum|MAXimum}

[SOURce:]CURRent:PROTection[:LEVel][:OVER]?

Parameter

Value PLZ164W 0.03 to 36.3 (Resolution: 10m, Default: 36.3)
PLZ164WA 0.03 to 36.3 (Resolution: 10m, Default: 36.3)
PLZ164WL 0.05 to 55.00 (Resolution: 50m, Default: 55.00)
PLZ164WH 0.01 to 9.07 (Resolution: 10m, Default: 9.07)
PLZ334W 0.06 to 72.6 (Resolution: 20m, Default: 72.6)
PLZ334WL 0.1 to 110.0 (Resolution: 20m, Default: 110.0)
PLZ334WH 0.01 to 18.15 (Resolution: 10m, Default: 18.15)
PLZ664WA 0.13 to 145.2 (Resolution: 10m, Default: 145.2)
PLZ1004W 0.20 to 220 (Resolution: 10m, Default: 220)
PLZ1004WH 0.01 to 55 (Resolution: 10m, Default: 55)
Unit A

Table shows the setting that is specified when *RST or *RCL is sent.

Response

Returns the overcurrent protection setting in the NR3 form in response to CURR:PROT?