Default Settings
Conditions after sending a *RST, fctory shipment, and at power-on
The following table shows how the PCR-M is set when the *RST command is executed, at the time of factory shipment, and when the power is turned on.
Setup item | Setting | Unit | |||
---|---|---|---|---|---|
*RST | *RCL | Factory default | At power-on | ||
OUTP:COUP | AC | Depends on the settings in the memory | AC | Setting immediately before turning the POWER switch off | -- |
VOLT:RANG | 135 | 135 | -- | ||
VOLT:RANGE:AUTO | 0/OFF | 0/OFF | -- | ||
VOLT | 0 | 0 | V | ||
VOLT:OFFS | 0 | 0 | V | ||
FREQ | 60 | 60 | HZ | ||
OUTP | 0/OFF | No change | 0/OFF | 0/OFF | -- |
CURR | PCR500M:5.25 PCR1000M:10.5 PCR2000M:21.0 PCR4000M:42.0 |
Depends on the settings in the memory | PCR500M:5.25 PCR1000M:10.5 PCR2000M:21.0 PCR4000M:42.0 |
Setting immediately before turning the POWER switch off | A |
CURR:OFFS | PCR500M:4.2 PCR1000M:8.4 PCR2000M:16.8 PCR4000M:33.6 |
PCR500M:4.2 PCR1000M:8.4 PCR2000M:16.8 PCR4000M:33.6 |
A | ||
CURR:PROT:STAT | 1/ON | 1/ON | -- | ||
FREQ:LIM | 500 | 500 | HZ | ||
FREQ:LIM:LOW | 40 | 40 | HZ | ||
VOLT:LIM | 275.0 | 275.0 | V | ||
VOLT:LIM:LOW | 0 | 0 | V | ||
VOLT:OFFS:LIM | 388.0 | 388.0 | V | ||
VOLT:OFFS:LIM:LOW | 0 | 0 | V | ||
TRIG:SOUR | BUS | BUS | BUS | BUS | -- |
FREQ:MODE | FIX | FIX | FIX | FIX | -- |
VOLT:MODE | -- | ||||
VOLT:OFFS:MODE | -- | ||||
FREQ:TRIG | 60 | Depends on the settings in the memory | 60 | Setting immediately before turning the POWER switch off | HZ |
VOLT:TRIG | 0 | 0 | V | ||
VOLT:OFFS:TRIG | 0 | 0 | V | ||
TRIG:SEQ2:SOUR TRIG:SYNC:SOUR |
No change | No change | IMM | -- | |
TRIG:SEQ2:PHAS TRIG:SYNC:PHAS |
No change | No change | 0 | DEG | |
TRIG:SEQ3:SOUR TRIG:ACQ:SOUR |
IMM | IMM | IMM | IMM | -- |
INIT:CONT:SEQ1 INIT:CONT:SEQ3 INIT:CONT:NAME TRAN INIT:CONT:NAME ACQ |
0/OFF | 0/OFF | 0/OFF | 0/OFF | -- |
DISP:AMM | RMS | Depends on the settings in the memory | RMS | Setting immediately before turning the POWER switch off | -- |
SYST:KLOC | No change | No change | 0/OFF | -- | |
SYST:CONF:BACK | 1/ON | 1/ON | -- | ||
SYST:CONF:TRAC | 0/OFF | Setting immediately before turning the POWER switch off | -- | ||
SENS:AVER | 0/OFF | -- | |||
SENS:HOLD | SHOR | -- |