INIT:CONT:SEQ3
Sets the sequence operation auto continue mode of the sequence 3 group.
When the sequence operation auto continue mode is turned on
If the trigger source parameter is set to IMM, the change or measurement starts immediately. When the operation is complete, a new change or measurement automatically starts.
If the parameter is set to BUS, the change or measurement starts after receiving a software trigger. When the operation is complete, the PCR-M/ PCR-LE waits for the next trigger.
When the sequence operation auto continue mode is turned off
The change or measurement currently in progress continues unless ABOR is sent. The operation does not automatically continue to the new change or measurement.
Command
INITiate:CONTinuous:SEQuence3 {ON|OFF|1|0}
INITiate:CONTinuous:SEQuence3?
Parameter
Value | ON(1) | Auto continue mode on |
OFF(0) | Auto continue mode off (default) |
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
Response
Response
Returns the sequence operation auto continue mode setting in NR1 format in response to INIT:CONT:SEQ3?