CURR:OFFS
Sets the DC current upper limit.
This is valid in DC mode (OUTP:COUP DC).
This is valid in DC mode and AC+DC mode (OUTP:COUP DC|ACDC).
Single-phase three-wire output or Three-phase output (optional): Sets the AC current upper limit of all phases at once.
Command
[SOURce:]CURRent:OFFSet[:IMMediate] {<numeric>|MINimum|MAXimum}
[SOURce:]CURRent:OFFSet[:IMMediate]? [{MINimum|MAXimum}]
Parameter
Value | PCR500LE: 0.35 to 3.85 (3.85 by default) PCR1000LE: 0.7 to 7.70 (7.70 by default) PCR2000LE: 1.4 to 15.40 (15.40 by default) PCR3000LE: 2.1 to 23.10 (23.10 by default) PCR4000LE: 2.80 to 30.80 (30.80 by default) PCR6000LE: 4.2 to 46.20 (46.20 by default) PCR9000LE: 6.3 to 69.30 (69.30 by default) |
Unit | A |
Value | PCR6000LE2: Single-phase output: 4.2 to 46.20 (46.20 by default) Single-phase three-wire output: 1.40 to 15.40 (15.40 by default) Three-phase output: 1.40 to 15.40 (15.40 by default) PCR9000LE2: Single-phase output: 6.30 to 69.30 (69.30 by default) Single-phase three-wire output: 2.10 to 23.10 (23.10 by default) Three-phase output: 2.10 to 23.10 (23.10 by default) PCR12000LE2: Single-phase output: 8.40 to 92.40 (92.40 by default) Single-phase three-wire output: 2.80 to 30.80 (30.80 by default) Three-phase output: 2.80 to 30.80 (30.80 by default) PCR18000LE2: Single-phase output: 12.60 to 138.60 (138.60 by default) Single-phase three-wire output: 4.20 to 46.20 (46.20 by default) Three-phase output: 4.20 to 46.20 (46.20 by default) PCR27000LE2: Single-phase output: 18.90 to 207.90 (207.90 by default) Single-phase three-wire output: 6.30 to 69.30 (69.30 by default) Three-phase output: 6.30 to 69.30 (69.30 by default) |
Unit | A |
Value | PCR500M: 0.1 to 4.2 (4.2 by defaul) PCR1000M: 0.2 to 8.4 (8.4 by default) PCR2000M: 0.4 to 16.8 (16.8 by default) PCR4000M: 0.8 to 33.6 (33.6 by default) |
Unit | A |
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
Response
Response
Returns the DC current upper limit in NR3 format in response to CURR:OFFS?.
In case of the single-phase three-wire output, CURR:OFFS? returns a comma-separated list of the DC current upper limit values <NR3> for U and V phases.
In case of the three-phase output, CURR:OFFS? returns a comma-separated list of the DC current upper limit values <NR3> for U, V, and W phases.