Command (Function)

Measurement

Second level node: SEQuence3 or ACQuire

The same operation is carried out regardless of the node that is used.

-> Tutorial

 

TRIG:SEQ3:SOUR
TRIG:ACQ:SOUR
Sequence 3: Trigger source.
INIT:SEQ3 Sequence 3: Starts the measurement trigger function.
INIT:NAME Starts the trigger function.
TRIG:SEQ3
TRIG:ACQ
Sequence 3: Software trigger.
INIT:CONT:SEQ3
INIT:CONT:NAME
Sets the auto continue mode.
SENS:AVER:CLE Clears the measured date.
FETC:VOLT Queries the measured voltage output without starting the measurement operation.
MEAS:VOLT Queries the measured voltage output after starting a new measurement operation.
FETC:CURR Queries the measured current output without starting the measurement operation.
MEAS:CURR Queries the measured current output after starting a new measurement operation.
ABOR Aborts the operation of all sequences.

>top