Regurated DC Power Supply PWX series

Communication Interface Manual


 

VOLT

Sets the voltage

When the PWX is set so that constant voltage is controlled externally (VOLT:EXT:SOUR VOLT), settings performed by this command are invalid.

Command

[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] {<numeric>|MINimum|MAXimum}

[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? {MINimum|MAXimum}

Parameter

Value A value in the range of the VOLT:LIM:LOW setting to 105 % of the rated output voltage (the default value is 0 V)
Unit V
A SCPI error (-222, "Data out of range") occurs if outside the range.

For the setting that is applied when *RST or *RCL is sent, see this table.

Response

Returns the voltage setting in NR3 format in response to VOLT?.

Overview

Setup

Overview of Messages

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Multi channel (VMCB)

Command for DCS

Regurated DC Power SUpplyRegurated DC Power Supply PWX Series