Regurated DC Power Supply PWX series

Communication Interface Manual


 

[-199,-100] | [-299,-200]
[-399,-300] | [-499,-400]
[100-]

A List of Errors

Command errors

An error in the range [ -199, -100 ] indicates that an IEEE 488.2 syntax error has been detected by the instrument’s parser. The occurrence of any error in this class shall cause the Command Error (bit 5) in the event status register to be set.

 

Error code Error message description
-100 Command error This is the generic syntax error.
-101 Invalid character A syntactic element contains a character that is invalid for that type.
-102 Syntax error An unrecognized command or data type was encountered.
-103 Invalid separator The parser was expecting a separator and encountered an illegal character.
-104 Data type error The parser recognized a data element different than one allowed.
-105 GET not allowed A Group Execute Trigger was received within a program message.
-108 Parameter not allowed More parameters were received than expected for the header.
-109 Missing parameter Fewer parameters were recieved than required for the header.
-110 Command header error An error was detected in the header.
-113 Undefined header The header is undefined for this device.
-120 Numeric data error This error is generated when parsing a data element that appears to be numeric, including the nondecimal numeric types.
-130 Suffix error This error is generated when parsing a suffix.
-131 Invalid suffix The suffix does not follow the syntax or the suffix is inappropriate for this device.
-134 Suffix too long The suffix contained more than 12 characters.
-138 Suffix not allowed A suffix was encountered after a numeric element which does not allow suffixes.
-140 Character data error This error is generated when parsing a character data element.
-141 Invalid character data Either the character data element contains an invalid character or the particular element received is not valid for the header.
-144 Character data too long The character data element contains more than twelve characters.
-148 Character data not allowed A legal character data element was encountered where prohibited by the device.
-150 String data error This error is generated when parsing a string data element.
-160 Block data error This error is generated when parsing a block data element.
-170 Expression error This error is generated when parsing an expression data element.
-180 Macro error This error is generated when defining a macro or executing a macro.

>top

Execution errors

An error in the range [-299, -200] indicates that an error has been detected by the instrument’s execution control block. The occurrence of any error in this class shall cause the Execution Error (bit 4) in the event status register to be set.

 

Error code Error message description
-200 Execution error (generic) This is the generic syntax error for devices that cannot detect more specific errors.
-203 Command protected Indicates that a legal password-protected program command or query could not be executed because the command was disabled.
-210 Trigger error Trigger error.
-211 Trigger ignored Indicates that a GET, *TRG, or triggering signal was received and recognized by the device but was ignored because of device timing considerations.
-213 Init ignored Indicates that a request for a measurement initiation was ignored as another measurement was already in progress.
-214 Trigger deadlock Indicates that the trigger source for the initiation of a measurement is set to GET and a subsequent measurement query was received.
-220 Parameter error Indicates that a program data element related error occurred.
-221 Settings conflict Indicates that a legal program data element was parsed but could not be executed due to the current device state.
-222 Data out of range Indicates that a legal program data element was parsed but could not be executed because the interpreted value was outside the legal range as defined by the device.
-223 Too much data Indicates that a legal program data element of block, expression, or string type was received that contained more data than the device could handle due to memory or related device-specific requirements.
-224 Illegal parameter value Used where an exact value, from a list of possible values, was expected.
-230 Data corrupt or stale Possibly invalid data; new reading started but not completed since last access.
-241 Hardware missing Indicates that a legal program command or query could not be executed because of missing device hardware.

>top

Device-specific errors

The occurrence of any error in this class shall cause the Device Dependent Error (bit 3) in the event status register to be set.

 

Error code Error message description
-350 Queue overflow A specific code entered into the queue in lieu of the code that caused the error. This code indicates that there is no room in the queue and an error occurred but was not recorded.
-360 Communication error Communication error when the flow control is turned off. This error applies when the RS232C interface is used.
-362 Framing error in program message Framing error. This error applies when the RS232C interface is used.
-363 Input buffer overrun Buffer overrun error. This error applies when the RS232C interface is used.
-364 Time out error Time out error. This error applies when the RS232C interface is used.

>top

Query errors

An error in the range [-499, -400] indicates that the output queue control of the instrument has detected a problem with the message exchange protocol described in IEEE 488.2, chapter 6. The occurrence of any error in this class shall cause the Query Error (bit 2) in the event status register to be set.

 

Error code Error message description
-400 Query error (generic) This is the generic query error for devices that cannot detect more specific errors.
-410 Query INTERRUPTED Received a new command before the response was read.
-420 Query UNTERMINATED The controller attempted to read the response after the device received an unsupported query or has not received a query.
The -100 "COMMAND ERROR" error and this error are stored in the error queue. The controller will time out.
-430 Query DEADLOCKED The error queue, input buffer, and output buffer are full when sending large binary data as a response, and the transmission timing is off.
-440 Query UNTERMINATED after indefinite response Received a separate query in semicolon-delimited format after a query that returns a response in an indefinite form.
(Example: A command such as the following. *IDN?;SYST:ERR?)

>top

Device-dependent errors

The occurrence of any error in this class shall cause the Device Dependent Error (bit 3) in the event status register to be set.

 

Error code Error message description
103 Operation denied while in PROTection state The operation was denied because a protection function has been activated.
141 CURR setting conflicts with CURR:PROT setting The set current conflicts with the set OCP.
142 CURR:PROT setting conflicts with CURR setting The set OCP conflicts with the set current.
151 VOLT setting conflicts with VOLT:PROT setting The set voltage conflicts with the set OVP.
152 VOLT:PROT setting conflicts with VOLT setting The set OVP conflicts with the set voltage.
153 VOLT setting conflicts with VOLT:LIM:LOW setting The set voltage conflicts with the set UVP.
154 VOLT:LIM:LOW setting conflicts with VOLT setting The set UVP conflicts with the set voltage.
155 Operation denied during ALARM condition The operation was denied because an alarm is occurring on the PWX.
901 EEPROM content error for CRITICAL data An error has occurred in a critical area. Contact your Kikusui agent or distributor.
902 EEPROM content error for CAL data An error has occurred in a calibration area. Contact your Kikusui agent or distributor.
903 EEPROM content error for CONFIG data An error has occurred in the configuration area. The PWX will be reset to its factory default settings. This error occurring multiple times indicates that the memory is damaged. Contact your Kikusui agent or distributor.
904 EEPROM content error for PANEL data An error has occurred in the panel memory area. The PWX will be reset to its factory default settings. This error occurring multiple times indicates that the memory is damaged. Contact your Kikusui agent or distributor.

>top

Overview

Setup

Overview of Messages

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Multi channel (VMCB)

Command for DCS

Regurated DC Power SUpplyRegurated DC Power Supply PWX Series