Regurated DC Power Supply PWX series

Communication Interface Manual


 

FAQ

The settings are not changed even when a trigger is sent.

Is the sequence in the WTG (Waiting for Trigger) state?

To change the setting using a trigger, set the trigger source to BUS and send the INIT command to set the sequence in the WTG state.

:TRIGger:SOURce BUS

:INITiate

If a trigger is sent in the WTG state, setting changes.

:TRIGger

->View details

The panel display is abnormal.

Are you communicating via LAN using the HiSLIP or SCPI-RAW protocol?

Using the HiSLIP or SCPI-RAW protocol may slow down the panel display updating. Even if this is the case, communication is operating normally.

If the delay is a concern, wait about 1 ms before sending the next command.

 