Regurated DC Power Supply PWX series

Communication Interface Manual


 

Overview of Messages

Default Conditions

Conditions after sending a *RST, *RCL, and at power-on

The following table shows how the PWX is set when *RST or *RCL is executed, at the time the PWX is shipped from the factory, and when the power is turned on.

 

Setup item Setting Unit Function
*RST *RCL Factory default At power-on
OUTP 0/OFF -- 0/OFF 0/OFF*1 -- OUTPUT on/off setting
VOLT 0 Depends on the settings in memory 0 Setting immediately before turning the POWER switch off -- Voltage setting
VOLT:LIM:LOW 0 0 V UVP setting
VOLT:LIM:AUTO ON ON ON -- Voltage limit setting
VOLT:PROT 112 % of the rated output voltage Depends on the settings in memory 112 % of the rated output voltage V OVP setting
CURR 105 % of the rated output current 105 % of the rated output current A Current setting
CURR:PROT 112 % of the rated output current 112 % of the rated output current A OCP setting
CURR:LIM:AUTO ON ON ON -- Current limit setting
VOLT:EXT:SOUR NONE NONE NONE -- Whether CV mode is being controlled externally
CURR:EXT:SOUR NONE NONE NONE -- Whether CC mode is being controlled externally
OUTP:EXT OFF OFF OFF -- Whether output is being turned on and off externally
VOLT:TRIG 0 The same value as VOLT 0 V Target voltage using a trigger
CURR:TRIG 105.0 % of the rated output current The same value as CURR 105.0 % of the rated output current A Target current using a trigger
TRIG:SOUR
TRIG:TRAN:SOUR
IMM IMM IMM IMM -- Trigger source

*1 The output may turn on at PWX power-on depending on the OUTP:PON setting.

>top

Overview

Setup

Overview of Messages

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Multi channel (VMCB)

Command for DCS

Regurated DC Power SUpplyRegurated DC Power Supply PWX Series
Communication Interface Manual