SENS:CURR:RANG

Sets the current range.

When you use this command to set the range, the auto current range feature (SENS:CURR:RANG:AUTO) is turned off.

Command

SENSe:CURRent:RANGe[:UPPer] {<numeric>|MINimum|MAXimum}

SENSe:CURRent:RANGe[:UPPer]? [{MINimum|MAXimum}]

Parameter

Value 5m, 10m, 20m, 50m, 100m, 200m, 500m, 1, 2, 5, 10, 20 (The default value is 20.)
Unit A

Table shows the setting that is specified when *RST is sent.

Response

Returns the current range in NR3 format in response to SENS:CURR:RANG?.