INTEG:TIM
Enables and disables the integration time.
Use INTEG:TIM:COUN to set the integration time.
During integration, only the queries are valid. If you send one of these commands to settings while the KPM1000 is performing integration, an error (-221, Settings conflict) will occur.
Command
INTEGrate:TIMer[:STATe] {ON|OFF|1|0}
INTEGrate:TIMer[:STATe]?
Parameter
Value | ON (1) | The integration time is enabled. |
OFF (0) | The integration time is disabled (this is the default value). |
Table shows the setting that is specified when *RST is sent.
Response
Returns whether the integration time is enabled in NR1 format in response to INTEG:TIM?.