Regurated DC Power Supply PWX series

Communication Interface Manual


 

CURR:PROT

Sets the overcurrent protection (OCP) value.

Command

[SOURce:]CURRent:PROTection[:LEVel] {<numeric>|MINimum|MAXimum}

[SOURce:]CURRent:PROTection[:LEVel]? {MINimum|MAXimum}

Parameter

Value Approximately 105 % to 110 % of the present current value (CURR; the default value is 111.5 % of the rated output current)
If you attempt to specify a value that is 105 % of the current value or less, an SCPI error (-222, "Data out of range") will occur.
If the PWX is set so that constant current is being controlled externally (CURR:EXT:SOUR VOLT), you can specify 10 % to 111.5 % of the rated output current.
Unit A

For the setting that is applied when *RST or *RCL is sent, see this table.

Response

Returns the OCP value in NR3 format in response to CURR:PROT?.

Overview

Setup

Overview of Messages

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register