SYST:CONF:POW{1|2|3}:EXP
Sets the anticipated maximum power (power-saving function) of each phase.
This command is valid in single-phase three-wire output or three-phase output (optional).
Turn the POWER switch off and on to enable the setting.
Master-slave parallel operation (option) is valid only when the connected PCR-LE Series are of the same model.
Command
Setting of U phase
SYSTem:CONFigure:POWer1:EXPected[:MAXimum] {<numeric>|MINimum|MAXimum}
SYSTem:CONFigure:POWer1:EXPected[:MAXimum]? [{MINimum|MAXimum}]
Setting of V phase
SYSTem:CONFigure:POWer2:EXPected[:MAXimum] {<numeric>|MINimum|MAXimum}
SYSTem:CONFigure:POWer2:EXPected[:MAXimum]? [{MINimum|MAXimum}]
Setting of W phase (Three-phase output only)
SYSTem:CONFigure:POWer3:EXPected[:MAXimum] {<numeric>|MINimum|MAXimum}
SYSTem:CONFigure:POWer3:EXPected[:MAXimum]? [{MINimum|MAXimum}]
Parameter
Value | PCR2000LE: 1000, 2000 (2000 by default) PCR3000LE: 1000, 2000, 3000 (3000 by default) PCR4000LE: 1000, 2000, 3000, 4000 (4000 by default) PCR6000LE: 1000, 2000, 3000, 4000, 5000, 6000 (6000 by default) PCR9000LE: 1000, 2000, 3000, 4000, 5000, 6000, 7000, 8000, 9000 (9000 by default) The setting range during master-slave parallel operation is “1000VA x the number of parallel units” to “rated output capacity x the number of parallel units.” The resolution is 1000VA x the number of parallel units. |
PCR6000LE2: 1000, 2000 (2000 by default) PCR9000LE2: 1000, 2000, 3000 (3000 by default) PCR12000LE2: 1000, 2000, 3000, 4000 (4000 by default) PCR18000LE2: 1000, 2000, 3000, 4000, 5000, 6000 (6000 by default) PCR27000LE2: 1000, 2000, 3000, 4000, 5000, 6000, 7000, 8000, 9000 (9000 by default) |
Unit | VA |
Response
Returns the maximum anticipated power of U phase in NR3 format in response to SYST:CONF:POW1:EXP?.
Returns the maximum anticipated power of V phase in NR3 format in response to SYST:CONF:POW2:EXP?.
Returns the maximum anticipated power of W phase in NR3 format in response to SYST:CONF:POW3:EXP?.