OPERation:INSTrument:ISUMmary{1|2|3} subregister
This is the subregister of bit 1 to bit 3 of the OPERation:INSTrument subregister. This is a 16-bit register that contains information about the normal operating conditions of the PCR-LE/PCR-LE2 Series
Of the parameters {1|2|3}, 1 represents U phase, 2 represents V phase, and 3 represents W phase.
For example, V phase is specified as "OPER:INST:ISUM2".
Bit | Bit weight | Bit name | Description |
---|---|---|---|
0 | 1 | NOT USED | -- |
1 | 2 | NOT USED | -- |
2 | 4 | NOT USED | -- |
3 | 8 | NOT USED | -- |
4 | 16 | MEASuring | Indicates whether measurement is in progress on the PCR-LE/PCR-LE2. |
5 | 32 | Waiting for TRIGger | Indicates whether measurement is in progress on the PCR-LE/PCR-LE2. |
6 | 64 | NOT USED | -- |
7 | 128 | BUSY | The period when the PCR-LE/PCR-LE2 is returning from sleep mode or when the output cannot be turned on after an OCP2 or OC has occurred |
8 | 256 | CV | CV output |
9 | 512 | NOT USED | -- |
10 | 1024 | SYNC | Indicates whether the synchronization function is in operation. |
11 | 2048 | SIMuration | Indicates whether a power line abnormality simulation is being performed. |
12 | 4096 | RISE | Indicates whether a soft start is being performed. |
13 | 8192 | INSTrument Summary | -- |
14 | 16384 | PROGram | Indicates whether a sequence is being performed. |
15 | 32768 | NOT USED | Always zero |