OPERation status register (PCR-LE/PCR-LE2)
The OPERation status register is a 16-bit register that contains conditions that are part of the PCR normal operations.
Bit | Bit weight | Bit name | Description |
---|---|---|---|
0 | 1 | NOT USED | -- |
1 | 2 | NOT USED | -- |
2 | 4 | NOT USED | -- |
3 | 8 | NOT USED | -- |
4 | 16 | MEASuring *1 | Indicates whether measurement is in progress on the PCR-LE/PCR-LE2. |
5 | 32 | Waiting for TRIGger *2 | Indicates whether the PCR-LE/PCR-LE2 is waiting for a trigger (TRIG) |
6 | 64 | NOT USED | -- |
7 | 128 | BUSY | The period when the PCR-LE/PCR-LE2 is returning from sleep mode or when the output cannot be turned on after an OCP2 or OC has occurred |
8 | 256 | CV | CV output |
9 | 512 | NOT USED | -- |
10 | 1024 | SYNC | Indicates whether the synchronization function is in operation. |
11 | 2048 | SIMuration | Indicates whether a power line abnormality simulation is being performed. |
12 | 4096 | RISE | Indicates whether a soft start is being performed. |
13 | 8192 | INSTrument Summary | Summary bit of OPERation:INSTrument sub register |
14 | 16384 | PROGram | Indicates whether a sequence is being performed. |
15 | 32768 | NOT USED | Always zero |
*1. A summary bit of the STAT:OPER:MEAS register. Further detailed information can be obtained by referring to this register.
*2. A summary bit of the STAT:OPER:TRIG register. Further detailed information can be obtained by referring to this register.