TRIG:SOUR
Sets the condition (trigger source) for actually measuring after the sequence 1 group receives INIT or INIT:NAME ACQ.
Command
TRIGger[:SEQuence[1]]:SOURce {IMMediate|BUS}
TRIGger[:SEQuence[1]]:SOURce?
TRIGger[:ACQuire]:SOURce {IMMediate|BUS}
TRIGger[:ACQuire]:SOURce?
Parameter
Value | IMM | Starts the setting immediately (Default) |
BUS | Wait for a software trigger (*TRG, TRIG, or IEEE488.1 get (Group Execute Trigger)) to change the setting |
For the setting that is applied when *RST is sent, see Table.
Response
Returns the trigger source as character data in response to TRIG:SOUR?.