READ:TIM:INTEG and FETC:TIM:INTEG
READ:TIM:INTEG? performs a measurement without changing the current settings and then queries the integration time.
FETC:TIM:INTEG? queries the integration time without starting a new measurement.
If you send the MEAS:TIM:INTEG? query while the KPM1000 is performing integration, an error (-221, Settings conflict) will occur.
Command
READ[:SCALar]:TIMe:INTEGrate?
FETCh[:SCALar]:TIMe:INTEGrate?
Response
Returns the integration time in NR3,NR3,NR3,NR3 format (a comma-separated list in which the data is in the order of hours, minutes, seconds, and milliseconds) in response to READ:TIM:INTEG? or FETC:TIM:INTEG?.
Unit: HR, MIN, S, MS