Setup
Interface Setup
GPIB (Option)
This interface valid only when the factory option GPIB interface board is installed.
GPIB connection
Use a standard IEEE488 cable to connect the KPM1000 to the computer.
GPIB configuration
1. Press I/F.
"SELECt" is displayed in the settings.
2. Press DISP D until "GPIb" is displayed.
3. Press DISP C until "AddrESS" is displayed.
4. Press the Δ∇ or DISP D to select the GPIB address (1 to 30).
5. Restart the KPM1000.
The settings are applied.
GPIB function
Function | Subset | Description |
---|---|---|
Source handshaking | SH1 | Full capability |
Acceptor handshaking | AH1 | Full capability |
Talker | T6 | Function available |
Listener | L4 | Function available |
Service request | SR1 | Full capability |
Remote local | RL1 | Full capability |
Parallel polling | PP0 | No capability |
Device clear | DC1 | Full capability |
Device trigger | DT1 | Full capability |
Controller | C0 | No capability |
Electrical interface | E1 | Open collector driver |
Service request
Service request and serial polling functions are implemented.