Electronic Load Setup
Sets the electronic load device.
Impedance measurement (current interrupt method)
Sets the load current of the PLZ-4W Series.
Command
SOURce:CURRent[:DC][:LEVel] {<numeric>|MIN|MAX|DEF}
SOURce:CURRent[:DC][:LEVel]? [{MIN|MAX}]
Parameter
Value: | 0 to the rated load current of the PLZ-4W (default: 0) |
Unit: | A |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the load current in the NR3 form in response to SOUR:CURR?.
Sets the load current of the external load unit.
Command
SOURce:CURRent[:DC]:EXTernal {<numeric>|MIN|MAX|DEF}
SOURce:CURRent[:DC]:EXTernal? [{MIN|MAX}]
Parameter
Value: | 0 to the value specified by Setup > Aux DC Load in the menu (default: 0) |
Unit: | A |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the load current of the external load unit in the NR3 form in response to SOUR:CURR:EXT.
Sets the range of the load current.
Command
SOURce:CURRent[:DC]:RANGe {LOW|MEDium|HIGH}
SOURce:CURRent[:DC] :RANGe?
Parameter
Value: | LOW, MEDium, or HIGH (default: HIGH) |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the range of the load current in response to SOUR:CURR:RANG?.
Sets the voltage of the PLZ-4W Series.
Command
SOURce:VOLTage[:DC][:LEVel] {<numeric>|MIN|MAX|DEF}
SOURce:VOLTage[:DC][:LEVel]? [{MIN|MAX}]
Parameter
Value: | 0 to 150 (default: 0) |
Unit: | V |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the voltage of the PLZ-4W Series in the NR3 form in response to SOUR:VOLT?.
Sets the voltage range of the PLZ-4W Series.
Command
SOURce:VOLTage[:DC]:RANGe {LOW|HIGH}
SOURce:VOLTage[:DC] :RANGe?
Parameter
Value: | LOW or HIGH (default: HIGH) |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the voltage range of the PLZ-4W Series in response to SOUR:VOLT:RANG?.
Sets the activation voltage of the undervoltage protection (UVP).
Command
SOURce:VOLTage[:DC]:PROTection:LOWer {<numeric>|MIN|MAX|DEF}
SOURce:VOLTage[:DC]:PROTection:LOWer? [{MIN|MAX}]
Parameter
Value: | -1 | Turns of the UVP function of the PLZ-4W (default). |
0 to 150 | ||
Unit: | V |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the activation voltage of the undervoltage protection (UVP) in the NR3 form in response to SOUR:VOLT:PROT:LOW?.
IMP:CURR:PULS:DEPT:RAT?.
Sets the activation power of the overpower protection (PLZ4W OPP).
Command
SOURce:POWer:PROTection:[Upper] {<numeric>|MIN|MAX|DEF}
SOURce:POWer:PROTection:[Upper]? [{MIN|MAX}]
Parameter
Value: | 0 to the rated load power of the PLZ-4W (default: rated load power) |
Unit: | W |
Table 4 shows the setting that is specified when *RST or *RCL is sent.
Response
Returns the activation power of the overpower protection (PLZ4W OPP) in the NR3 form in response to SOUR:POW:PROT?.
IMP:CURR:PULS:DEPT:RAT?.