Impedance measurement (current interrupt method)
Sets the pulse transition time of the current interrupt method.
Command
[SENSe:]IMPedance:CURRent:PULSe:TRANsition[:LEADing] {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:CURRent:PULSe:TRANsition[:LEADing]? [{MIN|MAX}]
Parameter
Value: | 0.000 01 to 0.01 (0.01E-3 to 10E-3): (default 0.000 01) |
Unit: | S (second) |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the pulse transition time of the current interrupt method in the NR3 form in response to IMP:CURR:PULS:TRAN?.
Sets the pulse width of the current interrupt method.
Command
[SENSe:]IMPedance:CURRent:PULSe:WIDTh {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:CURRent:PULSe:WIDTh? [{MIN|MAX}]
Parameter
Value: | 0.000 1 to 0.01 (0.1E-3 to 10E-3): (default: 0.01) |
Unit: | S (second) |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the pulse width of the current interrupt method in the NR3 form in response to IMP:CURR:PULS:WIDT?.
Sets the pulse depth of the current interrupt method as a percentage.
Command
[SENSe:]IMPedance:CURRent:PULSe:DEPTh:RATio {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:CURRent:PULSe:DEPTh:RATio? [{MIN|MAX}]
Parameter
Value: | 0 to 100 (default: 100) |
Unit: | PCT (%) |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the percentage of the pulse depth of the current interrupt method in the NR3 form in response to IMP:CURR:PULS:DEPT:RAT?.
Sets the pulse depth of the current interrupt method as a current value.
Command
[SENSe:]IMPedance:CURRent:PULSe:DEPTh[:LEVel] {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:CURRent:PULSe:DEPTh[:LEVel]? [{MIN|MAX}]
Parameter
Value: | 0 to the rated load current of the PLZ-4W (The default setting is the present DC load current setting (SOUR:CURR)) |
Unit: | A |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the pulse depth of the current interrupt method in the NR3 form in response to IMP:CURR:PULS:DEPT?.
IMP:CURR:PULS:DEPT:RAT?.
Sets the sampling start position (Position: time from interrupt start) of the current interrupt method.
Command
[SENSe:]IMPedance:SAMPling:POSition {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:SAMPling:POSition? [{MIN|MAX]}
Parameter
Value: | 0 to 0.009 99 (0 to 9.99E-3): (default: 0.000 06) |
Unit: | S (second) |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the sampling start position (Position: time from interrupt start) in the NR3 form in response to IMP:SAMP:POS?.
IMP:CURR:PULS:DEPT:RAT?.
Sets the sampling region of the current interrupt method.
Command
[SENSe:]IMPedance:SAMPling:REGion {<numeric>|MIN|MAX|DEF}
[SENSe:]IMPedance:SAMPling:REGion? [{MIN|MAX}]
Parameter
Value: | 0.000 002 to 0.000 51 (0.002E-3 to 0.51E-3): (default: 0.000 03) |
Unit: | S (second) |
Table 4 shows the setting that is specified when *RST, *RCL, or MEAS? is sent.
Response
Returns the sampling region of the current interrupt method in the NR3 form in response to IMP:SAMP:REG?.