AC power supply PCR-M series

Operation


Setting the Voltage Range

The voltage range is switched with the OUTPUT turned off. The selectable voltage ranges are 135 V, 270 V, and AUTO..

The RANGE key is disabled when the OUTPUT is turned on.

The AUTO setting automatically switches between 135 V range and 270 V range.

AUTO cannot be selected in AC+DC mode, EXT-AC mode, or EXT-DC mode.

23785.jpg

Voltage range and output voltage setting

Voltage Range

Output Voltage Setting

AC mode

DC mode

AC+DC mode

135 V range

0.0 V to 137.5 V

-194.0 V to +194.0 V

270 V range

0.0 V to 275.0 V

-388.0 V to +388.0 V

Voltage range and maximum output current

  

Voltage Range

Maximum Output Current

AC mode

DC mode

AC+DC mode

PCR500M

135 V range

5 A

4 A

270 V range

2.5 A

2 A

PCR1000M

135 V range

10 A

8 A

270 V range

5 A

4 A

PCR2000M

135 V range

20 A

16 A

270 V range

10 A

8 A

PCR4000M

135 V range

40 A

32 A

270 V range

20 A

16 A

1

Press the OUTPUT key to turn the OUTPUT off.

The LED above on the left of the OUTPUT key turns off.

2

Press the RANGE key (SHIFT+I) to select the voltage range.

The range switches among 135 V, 270 V, and AUTO each time the key is pressed. The LED corresponding to the range illuminates.

If the voltage is set above 137.5 V in the 270 V range or AUTO setting and the range is switched to 135 V, the output voltage is set to 0.0 V.