VOLT{1|2|3}:OFFS (PCR-LE/ PCR-LE2 only)
Sets the DC voltage of each phase.
This command is valid in DC mode or AC+DC mode (when the parameter to OUTP:COUP is DC|ACDC).
This command is valid in two-phase output(SYST:CONF:TPH:MODE ON) or three-phase output (optional).
If the EX05-PCR-LE is in use (optional) and the signal source is set to external signal source (FUNC:SOUR EXT), this command is invalid.
Command
U phase
[SOURce:]VOLTage1:OFFSet[:LEVel][:IMMediate][:AMPLitude] {<numeric>|MINimum|MAXimum}
[SOURce:]VOLTage1:OFFSet[:LEVel][:IMMediate][:AMPLitude]? [{MINimum|MAXimum}]
V phase
[SOURce:]VOLTage2:OFFSet[:LEVel][:IMMediate][:AMPLitude] {<numeric>|MINimum|MAXimum}
[SOURce:]VOLTage2:OFFSet[:LEVel][:IMMediate][:AMPLitude]? [{MINimum|MAXimum}]
W phase (Three-phase output only)
[SOURce:]VOLTage3:OFFSet[:LEVel][:IMMediate][:AMPLitude] {<numeric>|MINimum|MAXimum}
[SOURce:]VOLTage3:OFFSet[:LEVel][:IMMediate][:AMPLitude]? [{MINimum|MAXimum}]
Parameter
Value | DC mode, L range: -215.5 to 215.5 DC mode, H range: -431.0 to 431.0 AC+DC mode, L range: This can only be set when the AC and DC voltages are within the settable ranges defined by the voltage limits and the AC+DC waveform peak is between -215.5 V and 215.5 V. AC+DC mode, H range: This can only be set when the AC and DC voltages are within the settable ranges defined by the voltage limits and the AC+DC waveform peak is between -431 V and 431 V. A SCPI error (-222, "Data out of range") occurs if outside the range. The default value is 0. |
Unit | V |
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
For the setting that is applied when *RST is sent, see this table
Response
Response
Returns the DC voltage setting of U phase in NR3 format in response to VOLT1:OFFS?.
Returns the DC voltage setting of V phase in NR3 format in response to VOLT2:OFFS?.
Returns the DC voltage setting of W phase in NR3 format in response to VOLT3:OFFS?.